4–20 mA Calibration Procedure

Calibration ensures that the transmitter output accurately represents the process value across the entire 4–20 mA range.

Typical Calibration Loop Setup

24 VDC Transmitter mA Meter

Loop must always be measured in series.

Required Tools

Step 1 – Isolate the Instrument

Step 2 – Connect Calibration Setup

Step 3 – Zero Adjustment (4 mA)

Apply the lower range value (LRV).

Step 4 – Span Adjustment (20 mA)

Apply the upper range value (URV).

Step 5 – Intermediate Verification

Calibration Tolerance Calculation

If transmitter accuracy = ±0.5% of span:

Span = 16 mA (20 – 4)

0.5% of 16 mA = 0.08 mA

Acceptable 4 mA range = 3.92 to 4.08 mA
Acceptable 20 mA range = 19.92 to 20.08 mA

Common Field Mistakes

Field Rule

Never calibrate a transmitter before confirming correct loop wiring and stable power supply. Calibration does not fix wiring errors.

Related Blog Articles

Loop Check vs Calibration – What Most Get Wrong

Understand why calibration alone cannot confirm signal integrity.

5 Common 4–20 mA Mistakes in Field Work

Practical field errors observed during commissioning.