4–20 mA Calibration Procedure
Calibration ensures that the transmitter output accurately represents the process value across the entire 4–20 mA range.
Typical Calibration Loop Setup
Loop must always be measured in series.
Required Tools
- Calibrated multimeter or loop calibrator
- Process calibrator (pressure, temperature, etc.)
- 24 VDC loop power supply (if required)
- Manufacturer range data
Step 1 – Isolate the Instrument
- Inform control room and take necessary permits
- Isolate process (valves, manifolds)
- Put loop in manual or bypass mode
Step 2 – Connect Calibration Setup
- Break the loop at test terminals
- Insert multimeter or calibrator in series
- Confirm loop power is available
Step 3 – Zero Adjustment (4 mA)
Apply the lower range value (LRV).
- Expected output: 4.00 mA
- Adjust zero if required
Step 4 – Span Adjustment (20 mA)
Apply the upper range value (URV).
- Expected output: 20.00 mA
- Adjust span if required
Step 5 – Intermediate Verification
- Apply 25%, 50%, 75% input
- Confirm linear response
- Document as-found / as-left values
Calibration Tolerance Calculation
If transmitter accuracy = ±0.5% of span:
Span = 16 mA (20 – 4)
0.5% of 16 mA = 0.08 mA
Acceptable 4 mA range = 3.92 to 4.08 mA
Acceptable 20 mA range = 19.92 to 20.08 mA
Common Field Mistakes
- Calibrating without isolating impulse lines
- Measuring mA in parallel (wrong method)
- Ignoring power supply instability
- Not documenting as-found condition
- Forgetting damping configuration reset
Field Rule
Never calibrate a transmitter before confirming correct loop wiring and stable power supply. Calibration does not fix wiring errors.
Related Blog Articles
Loop Check vs Calibration – What Most Get Wrong
Understand why calibration alone cannot confirm signal integrity.
5 Common 4–20 mA Mistakes in Field Work
Practical field errors observed during commissioning.